PPS 1558-2

Project Plan 2026

Water quality — Application of inductively coupled plasma mass spectrometry (ICP-MS) — Part 2: Determination of selected elements including uranium isotopes

Sector: Environment
Project Method: Adoption with modification
Project Type: Update
International Classification for Standards (ICS)
Code Classification
13.060.50 Examination of water for chemical substances
Adopted Standard
Standards / Amendment / Corrigendum
1 ISO 17294-2:2023
Replaces
Will Replace
PS 1558-2-2019
Scope

This document specifies a method for the determination of the elements aluminium, antimony, arsenic, barium, beryllium, bismuth, boron, cadmium, caesium, calcium, cerium, chromium, cobalt, copper, dysprosium, erbium, gadolinium, gallium, germanium, gold, hafnium, holmium, indium, iridium, iron, lanthanum, lead, lithium, lutetium, magnesium, manganese, mercury, molybdenum, neodymium, nickel, palladium, phosphorus, platinum, potassium, praseodymium, rubidium, rhenium, rhodium, ruthenium, samarium, scandium, selenium, silver, sodium, strontium, terbium, tellurium, thorium, thallium, thulium, tin, titanium, tungsten, uranium and its isotopes, vanadium, yttrium, ytterbium, zinc and zirconium in water (e.g. drinking water, surface water, ground water, waste water and eluates).

Taking into account the specific and additionally occurring interferences, these elements can be determined in water and digests of water and sludge (e.g. digests of water as described in PS 1557-1 or PS 1557-2).

The working range depends on the matrix and the interferences encountered. In drinking water and relatively unpolluted waters, the limit of quantification (LOQ) lies between 0,002 μg/l and 1,0 μg/l for most elements (see Table 1). The working range typically covers concentrations between several ng/l and mg/l depending on the element and specified requirements.

The quantification limits of most elements are affected by blank contamination and depend predominantly on the laboratory air-handling facilities available on the purity of reagents and the cleanliness of glassware.

The lower limit of quantification is higher in cases where the determination suffers from interferences (see Clause 5) or memory effects (see PS 1558-1).

Elements other than those mentioned in the scope can also be determined according to this document provided that the user of the document is able to validate the method appropriately (e.g. interferences, sensitivity, repeatability, recovery).

Table 1 — Lower limits of quantification for unpolluted water

LOQ (a)
(µg/l)

Isotope often used

Element

LOQ (a)
(µg/l)

Isotope often used

Element

LOQ (a)
(µg/l)

Isotope often used

Element

0.1

102Ru

Ru

0.1

178Hf

Hf

0.5

107Ag

Ag

0.2

121Sb

Sb

0.05

202Hg

Hg

0.5

109Ag

0.1

201Hg

0.2

123Sb

0.1

165Ho

Ho

1

27Al

Al

5

45Sc

Sc

0.1

115In

In

0.1

75As(c)

As

1

77Se(c)

Se

0.1

193Ir

Ir

0.5

197Au

Au

0.1

78Se(c)

5

39K(c)

K

1

10B

B

1

82Se

0.1

139La

La

1

11B

0.1

147Sm

Sm

10

6Li

Li

3

137Ba

Ba

1

118Sn

Sn

1

7Li

0.5

138Ba

1

120Sn

0.1

175Lu

Lu

0.1

9Be

Be

0.5

86Sr

Sr

1

24Mg

Mg

0.5

209Bi

Bi

0.3

88Sr

10

25Mg

100

43Ca

Ca

0.1

159Tb

Tb

0.1

55Mn

Mn

50

44Ca

2

126Te

Te

0.5

95Mo

Mo

10

40Ca

0.1

232Th

Th

0.3

98Mo

0.1

111Cd

Cd

0.2

203Tl

Tl

10

23Na

Na

0.5

114Cd

0.1

205Tl

0.1

146Nd

Nd

0.1

140Ce

Ce

10

47Ti

Ti

1

48Ti

10

49Ti

0.1

169Tm

Tm

0.1

58Ni(c)

Ni

0.2

59Co

Co

0.1

238U

U

0.1

60Ni(c)

0.1

52Cr(c)

Cr

1.10−4

235U

5

31P

P

5

53Cr

1.10−5

234U

0.2

206Pb(b)

Pb

0.1

133Cs

Cs

0.1

51V(c)

V

0.2

207Pb(b)

0.1

63Cu

Cu

0.3

182W

W

0.1

208Pb(b)

0.1

65Cu

0.3

184W

0.5

108Pd

Pd

0.1

163Dy

Dy

0.1

89Y

Y

0.1

141Pr

Pr

0.1

166Er

Er

0.2

172Yb

Yb

0.5

195Pt

Pt

5

56Fe(c)

Fe

0.2

174Yb

0.1

85Rb

Rb

0.3

69Ga

Ga

1

64Zn

Zn

0.1

185Re

Re

0.3

71Ga

1

66Zn

0.1

187Re

0.1

157Gd

Gd

1

68Zn

0.1

103Rh

Rh

0.1

158Gd

0.2

90Zr

Zr

0.2

101Ru

Ru

0.3

74Ge

Ge

a Depending on the instrumentation, significantly lower limits can be achieved.

b Lead (Pb) is reported as the sum of the signal intensities of 206Pb, 207Pb and 208Pb.

c These limits are achieved by the use of a collision/reaction cell.